Pyrolysis GC-Orbitrap MS or GC-Orbitrap MS ((py-)GC-MS)

Description

Highly sensitive and accurate mass spectrometry (OrbiTrap MS) connected to a pyrolysis-GC combination for a versatile and dedicated application to characterise and quantify lignin in complex (insoluble) matrices as plant materials including biomass. A second GC unit connected to the OrbiTrap MS widens the number of applications of the system to quantify and identify unknown compounds, in particular fat and fat oxidation products, in complex mixtures.

The OrbiTrap MS is a powerful technique for unambiguous accurate mass determinations to 0.0001 amu (atomic mass unit (Da)) accuracy. The OrbiTrap has internal mass calibration and is equipped with multiple ionization possibilities (EI, VeV, CI) allowing detailed mass spectra for precise identification.

Technical Details

Pyrolysis unit:

  • FrontierLab Double-shot Pyrolyzer 3030D, with a cartridge heater (max: 900 °C), EGA temperature programming possibilities. Equipped with autosampler.

Gas chromatography:

  • GC1 (coupled to pyrolysis and the OrbiMS): Trace 1300; high resolution GC (125 °C/min till 450 °C).
  • GC2 (coupled to FID (instantConnect μFID‐detectionmodule) or OrbiMS); Trace 1300; high resolution GC (125 °C/min till 450 °C). With cold trap for large volume head space injections (TriPlus RSH multifunctional injectionrobot); with instantConnect BEST PTV‐LV backflush injectiemodule and Cold on Column injector.

Mass spectrometer:
Thermo Scientific Exactive series EI/VeV/CI quadrupole‐Orbitrap mass spectrometer:

  • mass-range 30‐>1000 amu
  • hyperquad mass filter with resolution from 0.4 amu
  • high resolution (60k) Orbitrap mass analyser with C‐trap ionfocussing
  • mass accuracy 0.0001 Da
  • sensitivity from 6 fg and linear dynamic range from >8 decades
  • ExtractaBrite EI/CI cassette with repeller and Rf MS‐pre‐filter
  • double digital gascontrol unit for CI‐gas

Applications

  • Pyrolysis-GC-MS for lignin analysis
  • GC-MS with cold trap facilities for head space injection, i.e. for fat and fat oxidation product analysis


Last edited by Oscar de Vos on 2023-05-31

Our expert(s)
Mirjam Kabel
Marie Hennebelle
Peter de Gijsel
Contact our expert(s)
Brand

Thermo Scientific 

Type

Trace 1300 GC with Exactive MS

Organisation

Food Chemistry
Shared Research Facilities

Subdivision

Food Chemistry



To explore the potential of nature to improve the quality of life.

This Shared Research Facilities equipment database offers researchers from universities, research institutes and companies, including start-ups, the opportunity to find research equipment with their relevant expert and to use it, generally on a pay-per-use basis. The equipment is available at different locations at Wageningen Campus or other (company) locations. If you want to share your own research devices or if you need more information, please contact us.